Sentinel V SC and RT Operation Manual
September
2017
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 227
PT – Built-In Tests
Purpose
Sends/displays results of Sentinel V Real-Time system diagnostic test.
Format
PTnnn
Range
nnn = 0, 3, 16 to 18, 21, 22 (PT0 = Help menu)
Recommended Setting. Use as needed.
DescriptionThese diagnostic tests check the major Sentinel V Real-Time modules and signal paths.
>pt0
Built In Tests
----------------
PT0 = Help
PT16 = Clock Test
PT17 = Compass Test
PT18 = Temperature Test
PT21 = Pressure Test
PT22 = FPGA Test>
PT16 = Clock Test
>pt16
RTC test................PASS
>
PT17 = Compass Test
>pt17
* indicates override value, selected by EZ command
Compass test........PASS [ 2.26, 1.39, -179.56 ]
PT18 = Temperature Test
>pt18
* indicates override value, selected by EZ command
Temperature test....PASS [ 25.335 ]
>
PT21 = Pressure Test
>pt21
* indicates override value, selected by EZ command
Pressure test.......PASS [ 117.00 ]
>
PT22 = FPGA Test
>pt22
FPGA test...........PASS