Operation
6.1
Environmental conditions
The lifetime of the electrochemical cartridges for toxic gas detector depends on the operating
environment related to temperature, humidity and high exposures.
One will take care to avoid long exposures to a hot and dry atmosphere (T°> 30°C and
RH<10%), and gas exposures 4 times higher to the scale.
Concerning to DGi-TT7-O type Y, avoid using the detector in close to the solvents with alcohol
or alcohol antiseptic products such as wipes and sanitizing gels.
The lifetime of the cartridges for semi-conductor toxic gas detector depends on the operating
environment related to certain compounds.
One will take care to avoid exposures to some vapored products as silicone. Moreover, an
exposure in area with a low rate of hygrometry for several hours can temporarily damage the
sensitivity. In order to retain stability of the factory span calibration, we recommand to leave in
place the protection cap with a desiccant packet on the front of the cartridge during storage,
shipping or any periods without power lasting more than 1 hour.
Don’t forget to remove the protection cap after power ON and before inserting
the eventual splash-guard.
Store the protection caps with the desiccant packet in a sealed container (zip-
lock bag). Dry the bag before using it again
In general, a dusty and humid atmosphere must be avoided. Indeed, a clogging of the gas
barriers is then possible, slowing down or stopping the detection process.
Summary of Contents for Everywhereyoulook DG-TT7-S
Page 32: ...MULTITOX DETECTOR OPERATING MANUAL 26 NOSP17668 Revision 03a ...
Page 34: ...MULTITOX DETECTOR OPERATING MANUAL 28 NOSP17668 Revision 03a ...
Page 47: ...MULTITOX DETECTOR OPERATING MANUAL NOSP17668 Revision 03a 41 ...
Page 54: ...MULTITOX DETECTOR OPERATING MANUAL 48 NOSP17668 Revision 03a ...
Page 58: ...MULTITOX DETECTOR OPERATING MANUAL 52 NOSP17668 Revision 03a ...
Page 60: ...MULTITOX DETECTOR OPERATING MANUAL 54 NOSP17668 Revision 03a ...
Page 69: ...MULTITOX DETECTOR OPERATING MANUAL NOSP17668 Revision 03a 63 10 Declaration of conformity ...
Page 70: ...MULTITOX DETECTOR OPERATING MANUAL 64 NOSP17668 Revision 03a ...
Page 71: ...MULTITOX DETECTOR OPERATING MANUAL NOSP17668 Revision 03a 65 ...