INTRODUCTION
6
Support
Online Documentation
Online Help is available by selecting Support > Dynamic Help from the
oscilloscope display menu bar. You can also select Support > OneTouch
Help for a demonstration of MAUI with OneTouch.
Teledyne LeCroy publishes a free Technical Library on its website.
Manuals, tutorials, application notes, white papers, and videos are
available to help you get the most out of your Teledyne LeCroy products.
The
WaveSurfer 510 Oscilloscopes Operator’s Manual
can be downloaded
from
teledynelecroy.com/support/techlib
under Manuals >
Oscilloscopes. This .PDF document contains more extensive procedures
for operating your oscilloscope than are found here.
The Datasheet published on the product page contains the detailed
product specifications.
Technical Support
Registered users can contact their local Teledyne LeCroy service center to
make Technical Support requests by phone or email. For a list of offices,
visit
teledynelecroy.com/support/contact
.
You can also submit Technical Support requests via the website at
teledynelecroy.com/support/techhelp
.
Two manual calibration routines can be used to recalibrate immediately
before sensitive measurements or when there is a significant change of
environmental characteristics that could affect the performance of the
ADCs:
Calibrate Current State
calibrates at the current vertical and horizontal
settings. This calibration takes 5 seconds but is valid for only this setup
at the current temperature ±2.5 °C.
Calibrate All
calibrates all possible combinations of vertical and
horizontal settings. This calibration is valid for the current temperature
and takes about four to five hours. Perform it when the ambient
temperature is stable from the time of calibration to the time of operation.
We recommend that you run Calibrate All when:
•
The instrument is operating outside the 19-30 °C range and the
operating temperature changes while outside that range. Calibrate
All is only valid for ± 2.5 °C of the last calibration temperature when
operating outside of the 19-30 °C range.
•
The instrument is moved to a new operating environment.
•
It has been six months since the previous manual calibration.
CAUTION.
Remove all inputs from the oscilloscope prior to manual
calibration.