Troubleshooting and Service
Teledyne API – Model T300/T300M CO Analyzer
278
QUESTION
ANSWER
How do I perform a leak check?
Section 10.3.3 provides leak check instructions.
How do I measure the sample
flow?
Sample flow is measured by attaching a calibrated rotameter, wet test
meter, or other flow-measuring device to the sample inlet port when the
instrument is operating. The sample flow should be 800 cm
3
/min
±
10%.
See Section 10.3.4.
How long does the IR source last?
Typical lifetime is about 2-3 years.
Can I automate the calibration of
my analyzer?
Any analyzer with zero/span valve or IZS option can be automatically
calibrated using the instrument’s AutoCal feature. The setup of this
option is located in Section 9.4.
Can I use the IZS option to
calibrate the analyzer?
Yes. However, whereas this may be acceptable for basic calibration
checks, the IZS option is not permitted as a calibration source in
applications following US EPA protocols.
To achieve highest accuracy, it is recommended to use cylinders of
calibrated span gases in combination with a zero air source.
Q:
What is the averaging time for
an T300/T300M?
A:
The default averaging time, optimized for ambient pollution
monitoring, is 150 seconds for stable concentrations and 10 seconds for
rapidly changing concentrations (see Section 12.5.1 for more
information).
However, it is adjustable over a range of 0.5 second to 200 seconds
(please contact Technical Support for more information).
11.8.
TECHNICAL ASSISTANCE
If this manual and its troubleshooting / repair sections do not solve your problems,
technical assistance may be obtained from:
Teledyne API Technical Support
9970 Carroll Canyon Road
San Diego, California 92131-1106 USA
Toll-free Phone:
800-324-5190
Phone:
+1 858-657-9800
Fax:
+1 858-657-9816
Email:
Website:
http://www.teledyne-api.com/
Before contacting Teledyne API Tech Support, please fill out the problem report form,
available online for electronic submission at http://www.teledyne-api.com.