menu from the Home Screen. This section provides information on the four status
outputs.
Figure 8-2. Status Outputs
Table 8-1. Status LED/Output Definitions
Status
Output #
Status LED
Label
Normal
Operating
State
Triggers
Critical
Warning?
1
Sensor OK
On
Reference < 125.0 mV OR
Reference > 3120.0mV OR
Measure > 3120.0mv
Yes
2
Invalid
Reading
Off
O3 Concentration too low
Units
O
3
Concentration
PPM
<
-30
WT%
<
-5
g/Nm3
<
-15
O3 Concentration > F.S. Range
No
3
Check Lamp Off
Reference < 250.0 mV OR
Measure > 3120.0mV
No
4
Pneumatic
Error
Off
Pressure < 9.0 psia
Pressure >18.0 .0 psia
Flow < 500 cc/min
Flow > 1100 cc/min
No