AFA4000/E
VAV Airflow Controller / p.7
2.4
Operating conditions
2.4.1
General operation
The Airflow Monitor and associated equipment have been designed and tested in accordance
with the International Electrotechnical Commissions (IEC) safety requirements. The Airflow
Monitor conforms to IEC61010-1 (Safety Requirements for electrical equipment for
measurement, control and laboratory use) as it applies to IEC Class 1 (earthed) appliances, and
therefore meets the requirements of EC directive 2014/30/EU.
Adjustment, maintenance or repair of the equipment whilst covers are open or the equipment is
in operation is only to be done by skilled persons who are aware of the potential hazards.
Make unsafe equipment inoperative.
Examples of unsafe equipment:
•
Appears damaged
•
Fails to perform correctly
•
Has undergone transport stresses
•
Has been stored incorrectly
2.4.2
Environmental conditions
Ensure the equipment is used under the following conditions:
•
Indoors
•
Temperature, 5°C to 40°C
•
Relative humidity below 80% for temperatures up to 31°C, decreasing linearly to 50% at
40°C
•
Electrical supply fluctuations not to 10% of nominal voltage
The protection provided by the equipment may be impaired if the environmental
conditions do not lie within these parameters.
Summary of Contents for AFA4000/E
Page 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Page 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Page 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Page 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Page 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Page 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Page 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Page 158: ...AFA4000 E VAV Airflow Controller p 158...
Page 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...