Cable Specifications
Mains cable
3-core, 0.7mm², 2m long
Light Curtain cable
4-core, 0.5mm², 2m long
Keypad cable
RJ45, 3 m long
Sash Position Sensor cable
4-core, 0.2mm², 2.5m long
Personnel Sensor cable
5-core, 0.2mm², 3m long
Motor cable
2-core, 0.5mm², 2m long
6.2
Auto Sash motor drive unit
A Motor Drive Unit is used to automatically raise and lower the fume hood sash.
Three types of drives are available:
•
Rack and pinion motor drive.
•
Chain and sprocket motor drive.
•
Wire and pulley motor drive.
The optimal choice of sash drive type depends on whether:
•
The installation is part of a new build or retro-fit.
•
The size and configuration of the fume hood.
•
The fume hoods location.
None of these options preclude manual operation of the sash.
6.2.1
Rack and pinion motor drive
The Rack and Pinion Motor Drive uses a rack fitted directly onto the sash frame and a motor
drive assembly fitted to a framework above the fume hood.
This drive system is ideal for retro-fit installations where there is no access to the sash wires or
counterweights and / or where the sash is fitted to a metal framework, for example
combination type sash fume hoods.
6.2.1.1
Installing the rack and pinion drive
1.
Raise the sash to the fully open position.
2.
Offer up the motor assembly and rack to ensure that there is a clearance of at least 1.2 m
(47.24") above the top of the sash.
Note: The rack is supplied in 1m (39.37") lengths
Summary of Contents for AFA4000/E
Page 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Page 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Page 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Page 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Page 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Page 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Page 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Page 158: ...AFA4000 E VAV Airflow Controller p 158...
Page 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...