AFA4000/2
Airflow Monitor / p.92
7.
Calibration
7.1
AFA4000/2 Airflow Monitor
The airflow monitor must be calibrated when it is first installed, or when required during
use.
When the unit is powered up the following sequence of events occur:
1. The 12 Vdc power is applied to the airflow sensor and a delay on timer is initiated.
2. The alarm then performs a self-test on the display and all indicators.
3. At the end of the delay period the unit performs one of two options:
a.
The unit goes into normal operating
Run
mode when the monitor calibration has
been previously completed.
b.
When the
unit has not been calibrated, the unit displays ‘Unit requires Set up
--
press Enter
to access Set up menu’. The
Set Up
menu allows the unit to be
calibrated or configured (password required).
Note: When the unit has been set with a
Fan Stop / Start
pushbutton, the monitor
can be set to Run mode before it is calibrated, to allow the Fan to be Started. To
Enter the menu from the Run screen, press and hold the
Mute
button for 5
seconds or until the main menu appears.
During the set-up all alarms and output relays are inhibited.
To calibrate the monitor:
1.
From the
Run Screen
press and hold the
Mute
button for 5 seconds until the
Main
Menu
is displayed.
Note: When you go from the main menu to the run screen, a note briefly appears
saying
‘calibration
required’.
2.
Use the
↑
/
↓
buttons to select
Set Up Monitor
, then
Calibration
.
3.
Enter the password (the factory default password is 0-0-0-0) and press
Mute
to
continue.
Note: The Monitor Set up Menu also includes Calibration and Password sub-menus
which are used by engineers to set up the system. If you enter any of the Monitor
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...