AFA4000/2
Airflow Monitor / p.165
15.1.5
Regulatory
Specification
AFA4000
Agency listings
CE RoHS
Hazardous area (remote airflow sensor) Gas group IIC Temp Class T6
with ATEX certified I.S. Zener Barriers
15.2
Auto Sash Controller
15.2.1
General
Auto Sash Controller
Options and features
Power Supply
1005
-240 Vac; 50/60 Hz; 3 A
•
2 m power lead with plug-in
connector
•
0.75 mm2, 300 V, 6 A rated
cable
Keypad
16*2 Backlit LCD display with
pushbuttons
•
UK single gang socket (85 mm
x 85 mm) version
•
US single gang socket (3" x 2")
version
Drive Motor
Assembly
PWM output DC motor and
clutch
•
Chain and sprocket shaft drive
•
Sash wire pulley drive
•
Rack and pinion sash direct
sash drive
Personnel
Sensor
PIR with background re-learn
function
Sash Position
Sensor
Steel wire sprung potentiometer
•
1 m length for bench type F/Cs
•
3 m length for walk-in type
F/Cs
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...