background image

Startup

The AFA1000/E must be field-calibrated once the 
room air supply and exhaust is proportionally 
balanced. When the unit is powered up, the 
following sequence of events occurs:
1.  The alarm performs a self-test of its 

functions, LEDs and audible alarm 
(approximately 3 seconds).

2.  At the end of the delay, the unit will do  

one of two things: 
a. If the controller has been calibrated, the 

unit enters normal operating mode (displays 
airflow velocity, LED’s and audible alarm 
enabled). 
b. If the controller has not been calibrated, 

the unit will display “ Requires set up, press 
Enter to continue”.

Installation & 

Calibration Guide

Airflow Controls

AFA1000/E MK2

Calibration

1.  Press Enter from the “Requires set up” screen or if the controller  

is in the Run screen Press and Hold the Enter button for 5 seconds 
until the Main Menu is displayed.

2.  Using the + / - buttons select SET UP, then select CONFIGURE, then 

enter the password (the factory default password is 0-0-0-0), press 
Enter to continue.

3.  Using the + / - buttons select ECON CONFIGURE, then select 

MANUAL / AUTO and press Enter, select MANUAL and press  
Enter to continue. 

4.  Open the sash to the normal operating height and measure the 

face velocity using a calibrated instrument. Using the + / - buttons 
adjust the Manual output so that the face velocity is equal to the 
design velocity, e.g. 0.50m/sec (100fpm) , then press Enter, then 
select DONE and press Enter again.

5.  Using the + / - buttons select SET UP, then select CALIBRATION, 

then enter the password (the factory default password is 0-0-0-0), 
press Enter to continue.

6.  With the sash open to the normal operating height again measure 

the face velocity using a calibrated instrument. Using the + / 
-buttons enter the measured face velocity then press Enter, the 
controller will then sample the airflow for 5 seconds.

7.  If the airflow sample is unstable the controller will display 

“Deviations too High”, follow the instructions to repeat the  
sample or quit the calibration.