![Tektronix WVR6020 Technical Reference Download Page 24](http://html1.mh-extra.com/html/tektronix/wvr6020/wvr6020_technical-reference_1077619024.webp)
Speci
fi
cations
Table 17: Ground closure remote / LTC time code input characteristics (cont.)
Characteristic
Performance requirement
Reference information
Ground Closure Input
signaling
TTL thresholds, 5 V max
input, –0.5 V min input. Pull
low to assert.
Has internal 10 k
Ω
pull-up to 5 V on each input. Direct or
encoded inputs must be asserted and stable for at least 150 ms
to be recognized reliably. Inputs that are stable for 75 ms or less
will never be recognized.
Hex
Binary Pins
15, 14, 13, 5
Direct mode
selection
Encoded
mode
selection
F
1111
none
No action
E
1110
Preset 1
CPS B
D
1101
Preset 2
CPS A
C
1100
SDI B
B
1011
Preset 3
SDI A
A
1010
Channel B
9
1001
Channel A
8
1000
7
0111
Preset 4
6
0110
Factory
Preset
5
0101
Preset 5
4
0100
Preset 4
3
0011
Preset 3
2
0010
Preset 2
1
0001
Preset 1
0
0000
unused
Ground Closure Output
characteristics
Open collector output,
30 V maximum, 100 mA
maximum, 300 mW
maximum.
Table 18: Power supply characteristics
Characteristic
Performance requirement
Reference information
Electrical Rating, typical
100-240 VAC ±10%, 50/60 Hz, 100 watts
maximum.
Continuous range from 90 to 264 VAC.
Supply Type
Single phase
Supply Connection, typical
Detachable cord set with lock.
Power Consumption, typical
< 90 VA (50 watts).
Fuse
Primary fuse value 3.15 AH (20 mm x
5 mm).
Not operator replaceable. Please refer to
quali
fi
ed personnel.
10
Waveform Rasterizers Speci
fi
cations and Performance Veri
fi
cation
Summary of Contents for WVR6020
Page 2: ......
Page 6: ......
Page 14: ...Preface viii Waveform Rasterizers Specifications and Performance Verification...