Audio Performance Veri
fi
cation
Audio Performance Veri
fi
cation
The tests in this section apply to instruments with audio options installed. The
tests will note when tests or their steps apply only to speci
fi
c audio options;
otherwise, they apply to all the audio options.
Required Equipment
Table 65: Required audio test equipment
Test equipment
Requirements
Example
Generates signals listed below.
Tektronix TG2000 with modules indicated
below.
Option HD:
1080i 59.94 100% color bars with two or
more groups of embedded audio
HDVG1 module for TG2000
SDI serial digital video test generator with
embedded audio
525/270 100% color bars with two or
more groups of embedded audio
DVG1 module for TG2000
Sine wave generator
Capable of 0.8 Vp-p at 6 MHz into a
50
Ω
load
Rohde & Schwarz SMT03 Opt. SM-B2
Test oscilloscope
Amplitude measurements of a 6 MHz
sine wave
Tektronix TDS7404B
Analog/Digital audio generator/analyzer
35 kHz to 96 kHz sample rate range, jitter
measurement per AES-3 (1997)
Rohde & Schwarz UPL06 Opt B22, B29
Precision 75
Ω
-to-50
Ω
adapter
1.5% impedance and attenuation
accuracy
Tektronix AMT75
75
Ω
coaxial cables
(2 required)
General purpose digital video
Male-to-male BNC connectors 3’ long
Belden 8281
50
Ω
coaxial cables
(2 required)
Male-to-male BNC connectors
36 inches long
Tektronix part number: 012-0482-00
75
Ω
Terminator
Male connector
Tektronix part number: 011-0055-02
Audio breakout cable
DB62 to XLR I/O required for Audio
Options AD, DD, DDE
Tektronix part number: 012-1688-00
DB37 to XLR I/O required for Audio
Option DA
Tektronix part number: 012-1658-01
Calculator
Logarithmic function
Dolby (R) digital audio generator
Generates Dolby D and Dolby E bit
streams.
Required for Options DD, and DDE.
Dolby Laboratories DM100
110
Waveform Rasterizers Speci
fi
cations and Performance Veri
fi
cation