To manually measure rise
time using variable gain
1.
Perform the initial setup for eye measurements. See
on page 191.
2.
Press and hold the
GAIN
button to view the Gain pop-up menu and select
Enable
for
VAR Gain
. Press the
GAIN
button again to dismiss the pop-up
menu.
3.
Use the
General
knob to resize the waveform to 10 major divisions. Position
the top and bottom of the waveform on graticule lines.
4.
Press the
CURSOR
button to display the measurement cursors.
5.
Position the first timing cursor at the crossing of the rising edge of the eye
waveform and the graticule line two divisions above the bottom of the
waveform. (See the following image.)
6.
Position the second time cursor at the crossing of the rising edge of the eye
waveform and the graticule line two divisions below the top of the waveform.
(See the following image.)
7.
Note the 20 – 80% rise time measurement displayed in the Delta time
readout. (See the following image.)
Figure 57: Measuring rise time using variable gain in the Eye display
Monitoring the SDI physical layer (WFM2300 only)
WFM2300 and WFM2200A Waveform Monitors User Manual
193
Summary of Contents for WFM2200A
Page 2: ......
Page 6: ......
Page 16: ...Table of Contents x WFM2300 and WFM2200A Waveform Monitors User Manual ...
Page 200: ...Functions 178 WFM2300 and WFM2200A Waveform Monitors User Manual ...
Page 208: ...Software upgrades 186 WFM2300 and WFM2200A Waveform Monitors User Manual ...
Page 242: ...ARIB content displays 220 WFM2300 and WFM2200A Waveform Monitors User Manual ...
Page 252: ...Audio monitoring 230 WFM2300 and WFM2200A Waveform Monitors User Manual ...
Page 270: ...Application examples 248 WFM2300 and WFM2200A Waveform Monitors User Manual ...