background image

The ground path for high frequency signal return currents is shown in the probe architecture figure. (See 

Figure 4: TDP7700

Series TriMode probe input architecture

 on page 4.) The probe ground path is continuous and along the following path:

• Extends from the DUT ground vias connections at the probe tip input
• Through the probe tip ground path
• Through the TekFlex connector ground spring contacts
• Along the main cable coaxial shields into the probe comp box
• Through the probe amplifier circuit board ground plane inside the probe comp box, and
• Through the TekVPI interface to the host oscilloscope ground

A TriMode tip requires a short DUT ground reference for making low noise single-ended probe measurements.

Although two ground via connections are available on the probe tip input, a single ground via wire connection is usually
adequate for making single-ended measurements of both the A and B input signals or a common-mode measurement, all of
which are ground referenced.

If only one single ended signal will be connected to the probe, the user has the choice of connecting the A input to the
signal and the B input to ground, or connecting the A input to the signal and the ground input of the probe to the ground
input of the DUT. In this situation, Tektronix recommends using A-B mode with the B input connected to ground. Reasons for
this recommendation include that with the B input left disconnected, there is a possibility of an interfering signal coupling into
the input of the probe and distorting the measured signal acquired on the A side.

A second reason for using A-B vs. A-ground is that it is often more convenient to connect the differential inputs of the probe
to a device and keep the wire lengths short. The ground connections of the probe are set back from the tip and may not be
as convenient to connect to a DUT with tightly spaced test points.

The optional accessory P77BRWSR is a variable-spacing probe tip which operates in Differential Input mode only. The
P77BRWSR Browser probe tip does not have a physical ground connection at the probe tip; but the differential
measurement process itself provides a high frequency virtual ground connection between the A and B signal input pins. A
low frequency ground connection is optionally available at a square-pin socket on the browser probe tip housing.

The comp box receives power and control signals from the oscilloscope through the TekVPI interface. The TekVPI interface
is an intelligent probe connection that includes many automated and manual control features. For example, there is a probe
S-parameter memory in the comp box that stores high frequency characterization data for the probe signal path.

The downloaded S-parameter data for a probe and attached probe tip is used by the oscilloscope to generate a DSP
correction filter for optimum measurement fidelity. Because this S-parameter data is unique and serialized for each probe
and probe tip, it only needs to be downloaded once to an oscilloscope.

Automatic probe tip type identification is another example of the intelligent operation of the TDP7700 Series probes. When
no probe tip is attached to the TekFlex connector of a TDP7700 Series probe, the TekFlex probe tip interface power is
disabled. When a probe tip is attached to the TekFlex connector, the attachment is detected by the probe and the probe
queries the probe tip memory to check for a valid identification header. If a valid probe tip type is verified by this TekFlex
data interface query, the probe tip power is enabled until the probe tip detachment is detected.

When probe tip power is enabled, an LED is activated on the probe tip, indicating that the TekFlex interface and attached
probe tip appear to be operating properly. If the probe tip type attached to the TekFlex connector is a browser probe tip, the
probe tip automatically detects the current tip spacing and communicates that information to the oscilloscope.

There are several calibrated tip spacing regions defined for the browser tip, which affect the frequency response of the
browser tip signal. Wider tip spacing tends to degrade the browser tip frequency response. The oscilloscope uses the
current tip spacing region information to select the optimum DSP correction filter for use with the browser tip measurements.

Input architecture with the P77C292MM SMA adapter

The P77C292MM adapter is a passive probe adapter; there are no active components within the adapter. It is intended to
connect the TDP7700 series probes to 50 Ω RF test points.

Theory of operation

TDP7700 Series TriMode™ Probes Technical Reference

6

Summary of Contents for Trimode TDP7700 Series

Page 1: ...TDP7700 Series TriMode Probes Technical Reference Register now Click the following link to protect your product www tek com register 077 1427 01...

Page 2: ...n in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc TriMode and...

Page 3: ...der in tip connection wire length 14 Using offset voltage to extend TekFlex solder in tip input voltage range 17 Making single ended measurements using the P77BRWSR differential probe tip 18 Temperatu...

Page 4: ...P77C292MM adapter 13 Figure 16 P77STFLXA solder tip attached with 10 mils 0 250 mm wires rise time 30 5 ps 15 Figure 17 P77STFLXA solder tip attached with 75 mils 1 90 mm wires rise time 27 9 ps 15 F...

Page 5: ...List of Tables Table 1 Probe tip attenuation factors 5 Table 2 Rise time and overshoot degradation versus lead length 17 List of Tables TDP7700 Series TriMode Probes Technical Reference v...

Page 6: ...and components when power is present Do not operate in wet damp conditions Be aware that condensation may occur if a unit is moved from a cold to a warm environment Do not operate in an explosive atm...

Page 7: ...ou read the marking WARNING indicates an injury hazard not immediately accessible as you read the marking CAUTION indicates a hazard to property including the product Symbols on the product The follow...

Page 8: ...Important safety information TDP7700 Series TriMode Probes Technical Reference viii...

Page 9: ...tight spaces a 2 92 mm adapter for connecting to fixtures and pre installed test points and a high bandwidth adjustable browser for debugging and non permanent connections Figure 1 TDP7700 Series Tri...

Page 10: ...signals from the probe comp box through the TekFlex connector interface Some of the wires carry bidirectional data such as queries and responses about the type of probe tip attached to the TekFlex co...

Page 11: ...struction provides better usability compared to the lower cost flex circuit tips The first time the probe tip is detected S parameter data for the probe tip is sent to the oscilloscope and probe tip s...

Page 12: ...s higher loading characteristic below 1 GHz has problems when taking measurements of signals such as MIPI that can be switched to an unterminated high impedance mode for low power operation The TDP770...

Page 13: ...SP correction filters that are unique to the probe tip The S parameter memory includes a unique data header used in automatically identifying the probe tip type and serial number The A and B signal ou...

Page 14: ...signal input pins A low frequency ground connection is optionally available at a square pin socket on the browser probe tip housing The comp box receives power and control signals from the oscillosco...

Page 15: ...ncy and convenience by enabling full differential signal characterization from a single soldered connection TekFlex active probe solder tips support the TriMode functionality of a TDP7700 Series probe...

Page 16: ...ovides the probe ground reference and is connected to the probe tip ground interconnect using the probe s cable coaxial shields Figure 6 TriMode input structure Operating voltages The TDP7700 Series p...

Page 17: ...e input signal dynamic range gain compression will result in increased linearity error and clipping The DC specifications of the probe are valid only over the input signal dynamic range The dynamic ra...

Page 18: ...esult the offset voltage control is no longer calibrated and will have twice the calibrated effect on the measured probe output The offset voltage control accessible from the attached oscilloscope fro...

Page 19: ...n by the probe amplifier maximizing the differential dynamic range Figure 9 Simplified Auto Tracking circuitry The A and B input signals are buffered by the active probe tip buffer amplifier and passe...

Page 20: ...is shown on the vertical axis and the B operating voltage window range of 5 25 V is shown on the horizontal axis The dynamic range is also shown on the figure as the smaller green square The TekFlex s...

Page 21: ...operating voltage for the TekFlex solder in tips the dynamic range square labeled Small Signal AC can be moved around within the limits of the operating voltage window by adjusting the A and B offset...

Page 22: ...nnecting to the probe input and output signal ports are de embedded to remove interconnect losses and signal path imperfections TekFlex probe tip signal performance is measured using a 4 port VNA meas...

Page 23: ...wing figures show the effect of input wire length variation on measured responses Note that these measurements in these images were made using a 20 Ghz P7720 probe Figure 16 P77STFLXA solder tip attac...

Page 24: ...older tip attached with 200 mils 5 08 mm wires rise time 34 ps Figure 20 P77STFLXA solder tip attached with 300 mils 7 62 mm wires rise time 42 8 ps The following table shows the rise time and oversho...

Page 25: ...solder in tips is 4 V to 4 V which is adjusted using the Probe Setup screen of the oscilloscope or the offset knobs on the oscilloscope front panel Using the offset voltage controls it is possible to...

Page 26: ...d Offset B settings are both adjusted to match the DC common mode voltage of the A and B input signals A B 2 The differential Offset voltage should be set manually to the center of the signal voltage...

Page 27: ...sonable to try to make a skew measurement by directly comparing the A and B signal timing Although it is possible to display both the single ended A and B signals that make up the differential signal...

Page 28: ...fferential amplifier used to make differential measurements that reject any voltage common to the inputs and amplifies any difference between the inputs Voltage that is common to both inputs is often...

Page 29: ...ower the input impedance of the probe relative to the source impedance the lower the CMRR for a given source impedance imbalance Differences in the source impedance driving the two inputs lowers the C...

Page 30: ...log S32 S31 B Mode A typical isolation plot for the TDP7700 Series probes using a TekFlex solder tip is shown in the following figure When the probe is used with TekFlex solder tips note that channel...

Page 31: ...nvironmental on page 27 Output zero typical 4 mV Linearity typical 1 Termination voltage range typical 2 5 mV Overload indicator range typical P77C292MM tip only The Overload indicator is active when...

Page 32: ...4 65 mV RMS 4 65 mV RMS 4 65 mV RMS TDP7706 4 1 mV RMS 4 1 mV RMS 4 1 mV RMS TDP7704 4 1 mV RMS 4 1 mV RMS 4 1 mV RMS System Noise Typical Probe A B mode C mode D mode TDP7710 3 6 mV RMS 2 8 mV RMS 3...

Page 33: ...10 V to 10 V SMA adapter 4 V to 4 V Input range typical Solder in tips Single ended Differential 2 5 Vpp 5 0 Vpp Browser Single ended Differential 6 0 Vpp 12 0 Vpp SMA adapter Single ended Differenti...

Page 34: ...Figure 28 P77STCABL SE impedance Figure 29 P77STFLXA differential impedance Figure 30 PST77FLXA SE impedance Specifications TDP7700 Series TriMode Probes Technical Reference 26...

Page 35: ...rating 0 C to 50 C 32 F to 122 F Non Operating 20 C to 60 C 4 F to 140 F Cable and solder in tips Operating 35 C to 85 C 31 F to 185 F Non Operating 35 C to 85 C 31 F to 185 F SMA adapter Operating 35...

Page 36: ...Nonoperating 5 to 90 RH Relative Humidity at up to 40 C 5 to 55 RH above 40 C up to 60 C non condensing Specifications TDP7700 Series TriMode Probes Technical Reference 28...

Page 37: ...to a channel on the oscilloscope The LED on the tip should light up after loading 2 Set the probe to Differential A B mode 3 Adjust the vertical resolution so the noise is on the screen and isn t cli...

Page 38: ...Mode rejection ratio 21 DMRR 21 DSP correction filtering 14 E error conditions LED indicators 2 I input network 18 input voltage 9 L LED indicators 2 O offset voltage 10 operating voltage window 12 op...

Reviews: