Reference
This chapter describes the menus and operating details associated with each
front-panel menu button or control.
Acquire
Push the Acquire button to set acquisition parameters.
Options
Settings
Comments
Sample
Use to acquire and accurately display most
waveforms; this is the default mode.
Peak Detect
Use to detect glitches and reduce the
possibility of aliasing.
Average
Use to reduce random or uncorrelated noise
in the signal display; the number of averages
is selectable.
Averages
4, 16, 64, 128
Select number of averages.
Key Points
If you probe a noisy square wave signal that contains intermittent, narrow glitches,
the waveform displayed will vary depending on the acquisition mode you choose.
Sample
Peak Detect
Average
Sample.
Use Sample acquisition mode to acquire 2500 points and display them at
the SEC/DIV setting. Sample mode is the default mode.
Sample acquisition intervals (2500)
• Sample points
Sample mode acquires a single sample point in each interval.
TPS2000B Series Digital Oscilloscope User Manual
93
Summary of Contents for TPS2012B
Page 1: ...xx TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual P071273302 071 2733 02...
Page 2: ......
Page 3: ...TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual xx www tektronix com 071 2733 02...
Page 8: ......
Page 22: ...Compliance information xiv TPS2000B Series Digital Oscilloscope User Manual...
Page 26: ...Preface xviii TPS2000B Series Digital Oscilloscope User Manual...
Page 40: ...Getting Started 14 TPS2000B Series Digital Oscilloscope User Manual...
Page 93: ...Math FFT TPS2000B Series Digital Oscilloscope User Manual 67...
Page 96: ...Math FFT 70 TPS2000B Series Digital Oscilloscope User Manual...
Page 170: ...Appendix D Cleaning 144 TPS2000B Series Digital Oscilloscope User Manual...