Operating information

The TPP1000 10X Passive Probe is a high impedance probe
with 10X attenuation that is designed for use with Tektronix
5 Series MSO and MDO3000, MDO/MSO/DPO4000B,
MSO/DPO5000 series ground-referenced oscilloscopes.

Connect the probe as shown in the illustrations below.

Compensating the Probe

You should compensate the probe after you attach it to an
oscilloscope for the

fi

rst time, or after you have changed the

probe tip cartridge.

1.

Connect the probe to an oscilloscope channel.

2.

Connect the probe tip and ground to the probe
compensation terminals on the oscilloscope.

NOTE.

When using the MMCX tip (206-0663-xx), the following

part numbers are required to connect the tip to the Probe
Comp connections: 131-9717-xx (1), 196-3434-xx (1),
and 206-0569-xx (2). See Connecting the MMCX tip to the
oscilloscope Probe Comp.

3.

On 5 Series MSO instruments:

a.

Double tap the Channel Badge for the connected
probe.

b.

Select the Probe Setup.

c.

Select Compensate Probe.

d.

Select Compensate Probe in the pop up window.

4.

On MSO/DPO5000 instruments:

a.

Select Vertical > Probe Cal....

b.

Select the tab of the channel for the connected probe.

c.

In the Calibration section, click Calibrate Probe.

5.

On MDO3000 and MDO/MSO/DPO4000B instruments:

Press the Channel Menu front panel button for the

channel that you connected the probe to.

Push the More button until Probe Setup is selected.

Push the Calibrate Probe button and follow the

on-screen instructions.

If the probe compensation fails, verify that the signal and
ground connections are secure at the Probe Comp connections.
Also check that the ground connection is secure at the probe
head, the rigid, MMCX or pogo tip is secured tightly in the
probe head, and the hook tip is securely connected to the tip.

Standard accessories

WARNING.

To avoid electric shock when using the probe or

accessories, keep

fi

ngers behind the

fi

nger guard of probe

body and accessories.

Item

Description
Probe tips – pogo (white) and
rigid (gray)

The white pogo tip is
pre-installed on the probe, and
is spring-loaded for compliant
testing of circuit boards.
Reorder Tektronix part numbers:
206-0610-xx (rigid tip)
206-0611-xx (pogo tip)

Insulator sleeve

Unscrew this sleeve to replace
the probe tips.
Reorder Tektronix part number
342-1194-xx

Hook tip

Press the hook tip onto the probe
tip and then clamp the hook onto
the circuit.
Rating: 300 V CAT II

Reorder Tektronix part number
013-0362-xx

Micro hook tip

Use this tip to access test points
in tight spaces. Press the hook
tip onto the probe tip and then
extend the pincers around the
circuit.
Rating: 300 V CAT II
Reorder Tektronix part number
013-0363-xx

Universal IC cap

Use this cap to prevent shorting
the probe tip between IC pins.
Press the cap on the probe tip
until it snaps on, and then spin
the cap to expose the probe tip
toward the IC lead.
Reorder Tektronix part number
013-0366-xx

Ground springs

To limit aberrations on high
frequency signals caused by
ground path inductance, bend
the spring to reach nearby
ground connections (<0.75 in,
long; <0.25 in, short).

Reorder Tektronix part numbers:
016-2028-xx (long, 2 ea.)
016-2034-xx (short, 2 ea.)

Ground lead, with alligator clip

Secure the lead to the probe
head ground and then to your
circuit ground.

Reorder Tektronix part number
196-3521-xx

Color bands

Use these bands to identify the
oscilloscope channel at the probe
head.
Reorder Tektronix part number
016-0633-xx (5 pairs)

Optional accessories

NOTE.

The accessories shown below are available for the

probes and are rated

30 V unless indicated otherwise.

Accessory

Part number

MMCX probe tip (gold)

1

206-0663-xx

MMCX to 0.1-inch
(2.54 mm) square pin
adapter (blue), 0.025-inch
(0.635 mm) sq. pins

131-9717-xx

MMCX to 0.062 inch
(1.57 mm) square
pin adapter (white),
0.016 - 0.018 inch
(0.4 - 0.46 mm) sq. pins

131-9677-xx

Electrical Y-Lead

196-3434-xx

MicroCKT
Test Tip

206-0569-xx

BNC to Tip Adapter,
Unterminated

013-0367-xx

Circuit Board Test
Point/PCB Adapter

016-2016-xx

Chassis-Mount Probe
Test Jack

131-4210-xx

6” Clip-on Ground Lead

196-3198-xx

12” Alligator
Ground Lead

196-3512-xx

Wire, spool,
32 AWG

020-3045-xx

DUT Interface pin kit
with (qty. 20) 0.018 inch
(0.46 mm) round
solder-in pins

020-3169-xx

Probe tip tripod support
with living hinge, 2 each

2

352-1170-xx

Solder aid for 0.062-inch
(1.57 mm) pitch square
pins, 0.016 - 0.018-inch
(0.4 - 0.46 mm) sq. pins

003-1946-xx

1

See

Connecting the MMCX tip to the oscilloscope Probe Comp

for connection

instructions.

2

The tripod probe tip support is pictured with the probe tip and adapter attached.

Interchanging the Probe Tip

WARNING.

To reduce the risk of shock, disconnect the

probe before changing the probe tips.

For optimal performance, do a probe compensation after the
tip has been replaced.

Connecting the MMCX tip to the oscilloscope Probe Comp

WARNING.

To reduce the risk of shock, disconnect the

probe before connecting accessories.

Item

Description
Connect the MMCX tip to
adapter

1

The MMCX adapter
(131-9717-xx) snaps onto
the MMCX tip.

Connect the Y–lead to MMCX
adapter

1

Push the Y–lead (196-3434-xx)
adapter into the MMCX adapter
until it is seated.
The signal input (red) connects
to the center of the MMCX
adapter.

Connect the MicroCKT test
tips to Y-lead

Push the Y-lead into the handles
of the MicroCKT test tips
(206-0569-xx).

Connect the MicroCKT test tips to Probe Comp terminals

Attach the two MicoCKT test tips to the Probe Comp terminals on the
ocsilloscope front panel.

1

MMCX tip, adapter, and Y-lead is limited to 30 vrms, 42 Vpk, 60 VDC, and

OVT <500 Vpk.

TPP1000

1 GHz 10X Passive Probe

Instructions

x

*P071280903*

071-2809-03