background image

xx

Operating Information

The TPP0500 & TPP1000 10X Passive Probes are
high impedance probes with 10X attenuation that are
designed for use with Tektronix DPO/MSO4000B &
DPO/MSO5000 series ground-referenced oscilloscopes.

Connect the probe as shown in the illustrations below.

Compensating the Probe

You should compensate the probe after you attach it to an
oscilloscope for the

fi

rst time, or after you have changed

the probe tip cartridge.

1.

Connect the probe to an oscilloscope channel.

2.

On MSO/DPO5000 instruments:

a.

Connect the probe to the probe compensation output

terminals on the oscilloscope front panel.

b.

Select Vertical > Probe Cal....

c.

Select the channel number tab of the channel that you

connected the probe to.

d.

In the Calibration section, click Calibrate Probe.

On MSO/DPO4000B instruments:

a.

Press the Channel Menu front panel button for the

channel that you connected the probe to.

b.

Push the More button until Probe Setup is selected.

c.

Push the Calibrate Probe button and follow the

on-screen instructions.

If the probe compensation fails, verify that the signal
and ground connections are secure at the Probe Comp
connections on the oscilloscope. Also check that the ground
connection is secure at the probe head, the rigid or pogo
tip is secured tightly in the probe head, and the hook tip is
securely connected to the tip.

Standard Accessories

WARNING.

To avoid electric shock when using the probe

or accessories, keep

fi

ngers behind the

fi

nger guard of

probe body and accessories.

Item

Description
Probe tips

The rigid tip (gray tip) is
pre-installed on the probe.
The pogo tip (white tip) is
spring-loaded for compliant
testing of circuit boards.
Reorder part numbers:
206-0610-xx (rigid tip)
206-0611-xx (pogo tip)

Insulator sleeve

Unscrew this sleeve to
replace the probe tips. (See
procedure on next page).
Reorder Tektronix part
number 342-1194-xx

Hook tip

Press the hook tip onto the
probe tip and then clamp
the hook onto the circuit.
Rating: 300 V CAT II

Reorder Tektronix part
number 013-0362-xx

Micro hook tip

Use this tip to access test
points in tight spaces. Press
the hook tip onto the probe
tip and then extend the
pincers around the circuit.
Rating: 300 V CAT II
Reorder Tektronix part
number 013-0363-xx

Universal IC cap

Use this cap to prevent
shorting the probe tip
between IC pins.
Press the cap on the probe
tip until it snaps on, and
then spin the cap to expose
the probe tip toward the IC
lead.
Reorder Tektronix part
number 013-0366-xx

Item

Description
Ground springs

To limit aberrations on high
frequency signals caused
by ground path inductance,
bend the spring to reach
nearby ground connections
(<0.75 in, long; <0.25 in,
short). Reorder Tektronix
part numbers:
016-2028-xx (long, 2 ea.)
016-2034-xx (short, 2 ea.)

Ground lead, with
alligator clip

Secure the lead to the probe
head ground and then to
your circuit ground.

Reorder Tektronix part
number 196-3521-xx

Color bands

Use these bands to identify
the oscilloscope channel at the
probe head.
Reorder Tektronix part number
016-0633-xx (5 pairs)

Optional Accessories

The accessories shown below are available for the probes
and are rated

30 V unless indicated otherwise.

Accessory

Part
number

MicroCKT Test Tip

206-0569-xx

BNC to Tip Adapter,
Unterminated

013-0367-xx

Circuit Board Test
Point/PCB Adapter

016-2016-xx

Chassis-Mount
Probe Test Jack

131-4210-xx

6” Clip-on
Ground Lead

196-3198-xx

12” Alligator
Ground Lead

196-3512-xx

Wire, spool,
32 AWG

020-3045-xx

TPP0500 & TPP1000

500 MHz and 1 GHz 10X Passive Probes

Instructions

x

*P071280901*

071-2809-01