Operating Basics
Configuring Resume Measurement
To configure Resume measurement, follow these steps:
1.
Select the High Speed measurement tab> More button> Resume
measurement. The device/HUB resumes the High-Speed operation, which is
indicated by the presence of High-speed SOF packets (with 400 mV
nominal amplitude) following the K state driven by the host controller. For
the Host, this is the time between the falling edge of D+ to the First SOF.
This should not exceed 3.0 ms.
2.
To access the Configure Measurements menu, go to Measurements>
Configure.
Figure 19: Configuring Resume measurement
3.
Set the input Signal Direction to either Upstream or Downstream.
4.
Select the channel source in the D+ and D– fields. Use the drop down arrow
in the D+, D– field to set the source. The available options are: Ch1- Ch4
and Ref1-Ref4.
5.
Select the Run button.
6.
To generate reports, select Utilities> Report Generator.
40
TDSUSB2 Universal Serial Bus Measurement Package
Summary of Contents for TDSUSB2
Page 24: ...Getting Started 10 TDSUSB2 Universal Serial Bus Measurement Package...
Page 32: ...Installation 18 TDSUSB2 Universal Serial Bus Measurement Package...
Page 61: ...Operating Basics TDSUSB2 Universal Serial Bus Measurement Package 47...
Page 76: ...Taking Measurements 62 TDSUSB2 Universal Serial Bus Measurement Package...
Page 90: ...Tutorial 76 TDSUSB2 Universal Serial Bus Measurement Package...
Page 160: ...Application Examples 146 TDSUSB2 Universal Serial Bus Measurement Package...
Page 230: ...Measurement Algorithms 216 TDSUSB2 Universal Serial Bus Measurement Package...
Page 250: ...Glossary 236 TDSUSB2 Universal Serial Bus Measurement Package...