Probing Principles
Maximum Allowable Pulse Amplitude
To determine the maximum allowable pulse
amplitude, do the following:
1.
Measure the pulse width at the 50% points.
2.
Divide the Ampere-second (or
Ampere-microsecond) speci
fi
cation
for the range setting of the TCP0150 probe
by the pulse width.
The quotient is the maximum allowable
pulse amplitude; the peak amplitude of the
measured pulse must be less than this
value.
For example, the TCP0150 probe has
a maximum Ampere-second product of
15000 A-
μ
s in the 150 A range setting. If a
pulse measured with the probe has a width of
40
μ
s, the maximum allowable peak current
would be 15000 A-
μ
s divided by 40
μ
s, or
375 A.