Installation
Degaussing the Probe
1.
After the probe is identi
fi
ed by the host
instrument:
The screen display prompts you to
degauss the probe
The multicolor Degauss/AutoZero
status LED on the probe
fl
ashes red
NOTE.
The DC gain and offset are not
guaranteed when this LED
fl
ashes red.
2.
With no conductor in the jaw, lock the
probe slide.
3.
Press the
Degauss/AutoZero
button on
the probe or in the Degauss window on the
host instrument.
4.
The multicolor Degauss/AutoZero status
LED glows green to indicate a successful
degauss routine was run, and that the
probe is in normal operating mode.