![Tektronix TCP0020 Instruction Manual Download Page 95](http://html1.mh-extra.com/html/tektronix/tcp0020/tcp0020_instruction-manual_1078185095.webp)
Performance Veri
fi
cation
Test Record
Probe Model/Serial Number:
Temperature:
Date of Calibration:
Certi
fi
cate Number:
RH %:
Technician:
Performance test
Test current
Expected output
Minimum
Incoming
Outgoing
Maximum
DC gain accuracy
±10.0 A
1.000 V
-3%
+3%
Rise time
~1 A
<7.0 ns
NA
7.0 ns
Bandwidth
NA
>50 MHz
50 MHz
NA
20 A Current Probes Instruction Manual
71