Controls and Indicators
Degauss/AutoZero Button
The
Degauss/AutoZero
button degausses
and nulls any DC offset in the probe.
To degauss the probe, do the following:
1.
Disconnect the probe from any conductors,
and then close and lock the slider.
2.
Press the
Degauss/AutoZero
button to
initiate the degauss routine.
After successfully completing the Degauss
routine, the
Degauss/AutoZero
LED on
the probe glows green.