Saving data
Saving data
The oscilloscope has internal memory locations where you can save instrument setups and waveforms. You can also save
screen images, setups, and waveforms to
fi
les on an external USB drives.
Saving screen images to a
fi
le
Use this procedure to save a screen image to a
fi
le on a USB drive. A screen image is a ‘picture’ of the oscilloscope screen.
Screen images are useful to include in reports or to compare with other screen images over time.
NOTE.
A screen image is different than waveform data. A screen image is a picture of the screen, and does not contain any
data about the signals shown in the image. Waveform data is a
fi
le that contains the numeric values of all the sampled points
in a single channel waveform. You cannot use a screen image
fi
le for further analysis, whereas you can load the data from a
waveform
fi
le into other oscilloscopes, or into PC-based analysis programs to perform advanced signal measurements.
1.
Install a USB drive.
2.
Set up the display to show the waveforms and other readouts that you want to save.
3.
Push the
Save/Recall
front-panel button.
4.
Push the
Action
side-menu button.
5.
Use the
Multipurpose
knob to select and click
Save Image
.
6.
Push the
File Format
side-menu button.
7.
Select and click a graphical
fi
le format (
BMP
,
JPG
, or
PNG
).
8.
Push the
Save
side-menu button to save the screen image to an automatically-named
fi
le at the top level of the USB drive.
You can also set the
File Save
button to automatically save an image
fi
le to the USB drive. (See page 70,
to USB with the Save File button
About saved image
fi
le formats
The image save function saves screen information to three common image formats.
66