background image

These two voltage ratings depend on the probe and your application. Refer to the Specifications section of
the manual for more information.

Warning: To prevent electrical shock, do not exceed the maximum measurement or maximum
floating voltage for the oscilloscope input BNC connector, probe tip, or probe reference lead.

Connect and disconnect
properly

Connect the probe output to the measurement product before connecting the probe to the circuit under test.
Connect the probe reference lead to the circuit under test before connecting the probe input. Disconnect the
probe input and the probe reference lead from the circuit under test before disconnecting the probe from the
measurement product.

Connect and disconnect
properly

De-energize the circuit under test before connecting or disconnecting the current probe.

Connect the probe reference lead to earth ground only.
Do not connect a current probe to any wire that carries voltages or frequencies above the current probe
voltage rating.

Inspect the probe and
accessories

Before each use, inspect probe and accessories for damage (cuts, tears, or defects in the probe body,
accessories, or cable jacket). Do not use if damaged.

Service safety summary

The Service safety summary section contains additional information required to safely perform service on the product. Only qualified
personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any
service procedures.

To avoid electric shock

Do not touch exposed connections.

Do not service alone

Do not perform internal service or adjustments of this product unless another person capable of rendering
first aid and resuscitation is present.

Disconnect power

To avoid electric shock, switch off the product power and disconnect the power cord from the mains power
before removing any covers or panels, or opening the case for servicing.

Use care when servicing with
power on

Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable),
and disconnect test leads before removing protective panels, soldering, or replacing components.

Verify safety after repair

Always recheck ground continuity and mains dielectric strength after performing a repair.

Terms in the manual

These terms may appear in this manual:

Warning: Warning statements identify conditions or practices that could result in injury or loss of life.

CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other property.

Terms on the product

These terms may appear on the product:

• DANGER indicates an injury hazard immediately accessible as you read the marking.
• WARNING indicates an injury hazard not immediately accessible as you read the marking.
• CAUTION indicates a hazard to property including the product.

Important safety information

TBS1000C Series Digital Storage Oscilloscope Service Manual

9

Summary of Contents for TBS1000C Series

Page 1: ...TBS1000C Series Digital Storage Oscilloscope Service Manual P077158401 077 1584 01...

Page 2: ......

Page 3: ...TBS1000C Series Digital Storage Oscilloscope Service Manual Register now Click the following link to protect your product www tek com register 077 1584 01...

Page 4: ...ued and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of...

Page 5: ...ion and cleaning 12 General care 12 Interior cleaning 12 Exterior cleaning 12 Exterior inspection 12 Interior inspection 13 Cleaning procedure interior 13 Lubrication 14 Adjustments 15 Replaceable par...

Page 6: ...t 13 Table 2 Internal inspection checklist 13 Table 3 Replaceable parts list 17 Table 4 Replaceable parts LCD Front panel Front case Power supply Main board and Rear case 18 List of Tables TBS1000C Se...

Page 7: ...and certified for the country of use Ground the product This product is grounded through the grounding conductor of the power cord To avoid electric shock the grounding conductor must be connected to...

Page 8: ...als before you clean the product Provide proper ventilation Refer to the installation instructions in the manual for details on installing the product so it has proper ventilation Slots and openings a...

Page 9: ...ice procedures Read this Service safety summary and the General safety summary before performing any service procedures To avoid electric shock Do not touch exposed connections Do not service alone Do...

Page 10: ...ut the nature of the potential hazards and any actions which have to be taken to avoid them This symbol may also be used to refer the user to ratings in the manual The following symbols may appear on...

Page 11: ...al available on the Tektronix Web site Go to www tektronix com manuals and search for the user manual for your product available in 11 languages Programmer Commands The oscilloscope programmer manual...

Page 12: ...and disconnect it from line voltage before performing any procedure that follows Exterior cleaning CAUTION Avoid the use of chemical cleaning agents which might damage the plastics used in this oscill...

Page 13: ...oscilloscope Table 2 Internal inspection checklist Item Inspect for Repair action Circuit boards Loose broken or corroded solder connections Burned circuit boards Burned broken or cracked circuit run...

Page 14: ...replaced as an assembly if significant moisture or debris become trapped inside Cotton tipped applicators and lint free rags moistened in isopropyl alcohol and or deionized water offer the best clean...

Page 15: ...test equipment signal sources and measuring equipment to verify the performance of the instrument To determine if an adjustment is required use the Performance Verification procedure in the TBS1000C S...

Page 16: ...ision level of the part For some parts in this manual you will find the letters XX in place of the revision level number When you order parts Tektronix will provide you with the most current part for...

Page 17: ...arated from the description by a colon Because of space limitations an item name may sometimes appear as incomplete Use the U S Federal Catalog handbook H6 1 for further item name identification Abbre...

Page 18: ...panel board 8 131 6417 XX 1 Contact Electrical 9 220 0265 XX 3 BNC nuts 10 210 1039 XX 3 BNC locking washer 11 441 2909 XX 1 Chassis ASSY 12 211 1273 XX 6 Screw Threaded 13 870 1704 XX 1 Main PCB ass...

Reviews: