Compliance information
Compliance information
This section lists the EMC (electromagnetic compliance), safety, and environmental standards with which the instrument
complies.
EMC compliance
EU EMC Directive
Meets intent of Directive 2014/30/EU for Electromagnetic Compatibility. Compliance was demonstrated to the following
speci
fi
cations as listed in the Of
fi
cial Journal of the European Communities:
EN 61326-1, EN 61326-2-1.
EMC requirements for electrical equipment for measurement, control, and laboratory
use.
1 2 3 4
CISPR 11. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2. Electrostatic discharge immunity
IEC 61000-4-3. RF electromagnetic
fi
eld immunity
5
IEC 61000-4-4. Electrical fast transient/burst immunity
IEC 61000-4-5. Power line surge immunity
IEC 61000-4-6. Conducted RF immunity
6
IEC 61000-4-8. Power frequency magnetic
fi
eld immunity test
IEC 61000-4-11. Voltage dips and interruptions immunity
EN 61000-3-2.
AC power line harmonic emissions
EN 61000-3-3.
Voltage changes,
fl
uctuations, and
fl
icker
European contact.
Mfr. Compliance Contact
Tektronix, Inc. PO Box 500, MS 19
‐
045
Beaverton, OR 97077, USA
www.tek.com
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
For compliance with the EMC standards listed here, high quality shielded interface cables that incorporate low impedance
connection between the cable shield and the connector shell should be used.
4
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
5
Residual spurious signals in the RF section can typically increase to -60 dBm at 80 MHz to 1 GHz and to -45 dBm at 1.4 GHz to
2.7 GHz when the instrument is subjected to electromagnetic interference per the IEC 61000-4-3 test.
6
Residual spurious signals in the RF section can typically increase to -55 dBm when the instrument is subjected to electromagnetic
interference per the IEC 61000-4-6 test.
RSA7100 Series Quick Start User Manual
vii