![Tektronix RSA6100A Series Help Manual Download Page 323](http://html1.mh-extra.com/html/tektronix/rsa6100a-series/rsa6100a-series_help-manual_1078169323.webp)
Amplitude Corrections
External Gain/Loss Correction Tab
Settings
RF Mixer Level
for signal at
Ref. Level
Selected Acquisition
Bandwidth
IF Gain/Loss
settings
Notes and Trade-offs
Best for Multiple
Displays
-25 dBm
Acquisition Bandwidth is
set to the value required
to service as many of the
open displays as possible.
Neutral
Swept Speed, Dynamic
Range, Noise Floor
Auto
-25 dBm
Driven by the
measurement bandwidth
of the selected display. If
span is greater than the
available acquisition BW,
unit sweeps using 40 MHz
acquisitions
Neutral
Swept Speed, Dynamic
Range, Noise Floor may
not be optimal
Minimize
Sweep Time
-25 dBm
Unit uses widest available
acquisition BW for spans
greater than maximum
acquisition BW.
Neutral
For units equipped with
Option 110, swept update
rate is approximately 3
times faster than Auto.
No effect in units without
Option 110.
Maximize
Dynamic Range
-20 dBm
Driven by span and
user-selected RBW
when a spectrum-type
display is selected,
or by measurement
bandwidth and
fi
lter
bandwidth when other
display types are selected.
For RBW < 30 kHz,
the high dynamic range
312.5 kHz path is used
regardless of selected
span or measurement
bandwidth. Sweeps
are made in 40 MHz
acquisitions.
IF attenuation
increased to
improve dynamic
range. This
causes higher
noise level.
Might result in very slow
sweep speeds in wide
spans when RBW <
30 kHz.
Minimize Noise
-25 dBm
Same as Auto
Set to minimize
noise level.
Might result in earlier IF
overload under some
signal conditions.
xxx
External Gain/Loss Correction Tab
The External Gain/Loss Correction tab allows you to apply a correction to a signal to compensate for the
use of external equipment, such as an ampli
fi
er or attenuator. You can specify whether the corrections on
this tab apply to all acquired data or only to a set group of displays.
RSA6100A Series Real-Time Spectrum Analyzers Help
311
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...