Analyzing Data > Analysis Settings
Spectrum Time Tab
Spectrum Time Tab
This tab controls whether the Spectrum display uses the same Analysis Time parameters as all the other
views or uses a different Offset and Length. These settings do not apply to the DPX Spectrum display
because it has no acquisition record or analysis time.
Settings
Description
Spectrum Time Mode
Independent
Causes the spectrum analysis views to use the settings unique to those displays.
Use Analysis Time settings
Causes the spectrum analysis views to use the settings on the Analysis Time tab.
Spectrum Time (only available when
Independent
is selected)
Spectrum Offset
Sets the beginning of Spectrum Time with respect to the selected time reference point
(selectable in the Analysis Time tab as either Acquisition Start or Trigger).
Spectrum Length
The amount of data, in terms of time, from which spectrum traces are computed.
Auto
When enabled, causes the instrument to set the Spectrum Length value based on the
RBW setting.
Actual
This is a displayed value, not a setting. It is the Spectrum Length (time) being used
by the analyzer; this value may not match the Spectrum Length requested (in manual
mode). The actual spectrum length is always an integer multiple of the time needed to
support the RBW value.
xxx
Frequency Tab
The Frequency tab speci
fi
es two frequency values: the Measurement Frequency and the Spectrum Center
Frequency. The Measurement Frequency is the frequency at which most displays take measurements. The
Spectrum Center Frequency is the center frequency used by the Spectrum, DPX Spectrum, Spectrogram
and Time Overview displays.
RSA6100A Series Real-Time Spectrum Analyzers Help
299
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...