![Tektronix RSA6100A Series Help Manual Download Page 229](http://html1.mh-extra.com/html/tektronix/rsa6100a-series/rsa6100a-series_help-manual_1078169229.webp)
Pulsed RF > Common Controls for Pulsed RF Displays
De
fi
ne Tab
Setting
Description
Apply ampl corrections
Enable/disable corrections that remove errors due to the window function and to
the time offset of the side lobe.
Keep-out time (+/-)
The Keep-out time speci
fi
es a region that is ignored when the trace is analyzed for
side lobes. The setting de
fi
nes a region to the left and to the right of the center
of the main lobe. Lobes that fall within this time region are not eligible to be the
“highest side lobe”.
Time method
Speci
fi
es how measurement parameters are determined. The choices are
Absolute and Relative.
Absolute time method
Ref (R)
Ref speci
fi
es the level along the pulse rising edge that is de
fi
ned as the point
from which Start time is counted. Choices: 50%, 60%, 70%, 80%, 90% 100%;
Default: 100%
Start (S)
Start speci
fi
es the time the instrument waits after the rising-edge reference before
starting to measure the pulse. The Start time is measured from the point along the
pulse rising edge speci
fi
ed by the Ref setting. Start Range: ±100 ms; Resolution:
3 digits; inc/dec small: 1 ns, large: 1,2,5,10...; Default: 0
Length (L)
Amount of time that should be included in the measurement. The measurement
time begins at the Start point and continues for the amount of time speci
fi
ed by
Length.
Relative time method
In the Absolute time method, Length speci
fi
es the period of time that is used for
pulse measurements.
Length
In the Relative time method, Length speci
fi
es the percentage of the top of the pulse
that is used for measurements. The instrument automatically determines the pulse
top. The measurement length is centered within the pulse ON time.
xxx
Freq Diff, Phase Diff
Setting
Description
Measure point
Speci
fi
es the period in time after the 50% rising edge at which frequency and
phase difference measurements are made.
xxx
RSA6100A Series Real-Time Spectrum Analyzers Help