RF Measurements > Settling Time Displays
Settling Time Displays
Settling Time Displays
The Frequency Settling Time measurement is used to measure the frequency settling time of
frequency-agile oscillators and subsystems. The Phase Settling Time measurement is used to measure
the phase settling time of phase-agile oscillators and subsystems. Automated measurements can reduce
user-to-user measurement variations, improving repeatability and measurement con
fi
dence and saving
time in gathering results. To measure phase settling, a phase reference is required. This means that the
measurement tool must be frequency- referenced to the device under test, or must act as the frequency
reference to the device under test. To lock the reference to an external source, go to Setup > Con
fi
gure
In/Out > Frequency Reference and choose
External
. Values from 1 MHz to 25 MHz in 1 MHz steps
can be used as an external reference.
To show a Settling Time display:
1.
Click the
Displays
button or select
Setup
>
Displays
.
2.
From the Folders box, select
RF Measurements
.
3.
Double-click the
Freq Settling
icon or the
Phase Settling
icon in the
Available displays
box. This
adds the selected display to the
Selected displays
box (and removes it from the Available displays box).
4.
Click the
OK
button.
Frequency settling time display
182
RSA6100A Series Real-Time Spectrum Analyzers Help
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...