RF Measurements > Spurious
Ranges Tab
Ranges Tab
Use the Ranges tab to specify the parameters that control the Spurious measurement. From the Ranges tab,
you specify the start and stop frequencies for ranges, whether or not a range is enabled, the parameters
that specify what constitutes a spur, measurement
fi
lter shape and bandwidth, and the detection method
used for processing signals.
Setting
Description
Range
Selects the frequency range to which the settings on this tab are applied.
Start
Readout of the start frequency for the selected range.
Stop
Readout of the stop frequency for the selected range.
Edit
Click this button to edit the Spurious Frequency Plan, which de
fi
nes the range frequency
values and whether or not a signal is measured in a range (enabled/disabled).
Enabled
Speci
fi
es whether or not measurements are taken in the speci
fi
ed range.
Filter shape
Speci
fi
es the
fi
lter shape used for the Spurious measurement.
BW
Speci
fi
es the bandwidth used for the selected
fi
lter shape.
Auto
Sets the BW automatically. If CISPR is selected for Filter shape, this control is disabled.
Detection
Sets the processing method used for compressing excess intermediate data into the
desired number of trace points.
Threshold
Speci
fi
es the level that must be exceeded for a signal peak to be recognized as a spur. A
signal peak must also exceed the Excursion setting to be considered a spur.
Excursion
Speci
fi
es the peak to peak magnitude that must be exceeded for a signal peak to be
recognized as a spur. A signal transition must also exceed the Threshold setting to be
considered a spur.
VBW
Adjusts the VBW (Video Bandwidth) value. VBW Maximum: RBW current value; VBW
Minimum: 1/10,000 RBW setting. Disabled when Filter shape is set to CISPR.
Review table
Click to display the Ranges table. The Ranges Table shows the enabled ranges, and all
the settings speci
fi
ed on the Ranges and Limits tabs associated with the enabled ranges.
Save table
Click to save the current Range table to a
fi
le.
Load table
Click to load a saved Range table from a
fi
le.
xxx
162
RSA6100A Series Real-Time Spectrum Analyzers Help
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...