![Tektronix RSA3303B User Manual Download Page 22](http://html1.mh-extra.com/html/tektronix/rsa3303b/rsa3303b_user-manual_1078145022.webp)
Installation
Figure 8: Setup display
c.
Using the general purpose knob, change the span setting as listed below
and check that the RBW is displayed correctly.
Table 1: Span and RBW
Span
RBW
36 MHz (RSA3408B)
100 kHz
15 MHz
80 kHz
5 MHz
20 kHz
100 kHz
500 Hz
1 kHz
20 Hz
d.
Using the numeric keypad, set the span back to 36 MHz (RSA3408B) or
15 MHz (RSA3300B). (For example, press
3
→
6
→
MHz
, in that order,
on the keypad.)
5.
Check the reference level:
a.
Press the
AMPLITUDE
key on the front panel.
b.
Make sure that the reference level is set to 0 dBm with the
Ref Level
side
key. Check that 0 dBm is displayed on the upper left side of the graticule.
(See Figure 9.)
c.
Use the general purpose knob to set the reference level to -30 dBm.
d.
Con
fi
rm that A/D OVERFLOW is indicated in the red box at the top
center of the screen. Make sure that -30 dBm is displayed on the upper
left side of the graticule and that the spectrum waveform is distorted.
(See Figure 9.)
10
RSA3303B, RSA3308B, and RSA3408B User Manual
Summary of Contents for RSA3303B
Page 4: ......
Page 10: ...General Safety Summary vi RSA3303B RSA3308B and RSA3408B User Manual...
Page 16: ...Installation Figure 1 Setting up the stand 4 RSA3303B RSA3308B and RSA3408B User Manual...
Page 32: ...Calibration 20 RSA3303B RSA3308B and RSA3408B User Manual...
Page 48: ...Functional Overview 36 RSA3303B RSA3308B and RSA3408B User Manual...
Page 78: ...Appendix A Specifications 66 RSA3303B RSA3308B and RSA3408B User Manual...