Getting started
Ventilation
CAUTION.
Observe the following precautions to maintain proper ventilation:
Do not block any cooling vents.
Do not position any devices adjacent to the instrument that force air (heated or
unheated) into or onto the instrument’s surfaces or cooling vents. This additional
air
fl
ow could compromise accuracy performance.
When rack mounting the instrument, ensure there is adequate air
fl
ow around the
instrument rear and sides to ensure proper cooling. Adequate air
fl
ow enables air
temperatures within approximately one inch of the instrument surfaces to remain
within speci
fi
ed limits under all operating conditions.
Environmental ratings
The following table describes the maximum operating environment ratings for
your instrument.
Table 2: Maximum operating environment ratings
Feature
Description
Temperature
0 to 40 °C (32 to 104 °F)
Humidity
80% for temperatures up to 31 °C (88 °F) decreasing linearly
to 50% at 40 °C (104 °F)
Altitude
2000 m (6562 ft.)
Front panel controls and connectors
The front panel of the PSPL2600 is shown in the following
fi
gure. The Power
switch is located on the front left side of the instrument. The power indicator
adjacent to the switch lights up when the instrument is on.
Figure 1: PSPL2600C front panel controls and connectors
2
Tektronix PSPL2600C Pulse Generator User Manual