Functional Check and Calibration
TriMode Probe DC Calibration
After you perform a functional check of the probe, run a probe calibration routine on each channel that you use. The probe
calibration operation minimizes measurement errors by optimizing the DC gain and offset of the probe. Individual calibration
constants are stored for all TriMode settings, on each probe, on each channel.
CAUTION.
To avoid ESD damage to the probe, always use an antistatic wrist strap (provided with your probe), and work at
a static-approved workstation when you handle the probe.
Table 3: Required Equipment for Probe Calibration
Item description
Performance requirement
Recommended example
1
Oscilloscope
TekConnect Interface
Tektronix DPO73304DX, DPO72504DX
Test
fi
xture
Probe DC calibration
fi
xture
067-3259-xx
2
Adapter (any one of the adapters listed
can be used)
Interconnect between probe head and
probe calibration
fi
xture
P76CA-292C
P76CA-292 and two SMA cables
P76CA-SMP
P76TA
Coaxial cable
BNC, 50
Ω
, male-to-male
012-0208-xx
2
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers.
2
Standard accessory included with probe. (See page 17.)
P7600 Series TriMode Probes Quick Start User Manual
15
Summary of Contents for P7625
Page 1: ...x P7600 Series TriMode Probes ZZZ Quick Start User Manual P071302601 071 3026 01...
Page 2: ......
Page 3: ...P7600 Series TriMode Probes ZZZ Quick Start User Manual xx www tektronix com 071 3026 01...
Page 6: ......
Page 14: ...Preface viii P7600 Series TriMode Probes Quick Start User Manual...
Page 86: ...Maintenance 72 P7600 Series TriMode Probes Quick Start User Manual...