Functional Check and Calibration
Functional Check and Calibration
After you connect the probe to the oscilloscope, perform a functional check using the procedure below.
Functional Check
This procedure checks the four TriMode settings on the probe, using the FAST EDGE connection on the front panel of the
oscilloscope. The A-B (differential mode) is set up and veri
fi
ed
fi
rst, and then the remaining input modes are checked and
compared to the differential mode measurement.
Table 2: Required Equipment
Item description
Performance requirement
Recommended example
Oscilloscope
TekConnect Interface
Tektronix DPO73304DX or DPO72504DX
Probe adapter
P76CA-xxx or P76TA probe
adapter
P76CA-292C (connects directly to oscilloscope FAST EDGE connector)
P76CA-292 (requires a coaxial cable)
P76CA-SMP (requires an SMP-to-SMA adapter)
P76TA (requires a coaxial cable and the calibration
fi
xture listed below)
Coaxial cable
SMA, 50
Ω
, male-to-male
Tektronix part number 174-1120-xx
Adapter
SMP-to-SMA
Fairview Microwave part number SM8810
Probe calibration
fi
xture
Interconnect between P76TA
adapter and oscilloscope
Tektronix part number 067-1821-xx
P7600 Series TriMode Probes Quick Start User Manual
11
Summary of Contents for P7625
Page 1: ...x P7600 Series TriMode Probes ZZZ Quick Start User Manual P071302601 071 3026 01...
Page 2: ......
Page 3: ...P7600 Series TriMode Probes ZZZ Quick Start User Manual xx www tektronix com 071 3026 01...
Page 6: ......
Page 14: ...Preface viii P7600 Series TriMode Probes Quick Start User Manual...
Page 86: ...Maintenance 72 P7600 Series TriMode Probes Quick Start User Manual...