P5122 100X Probe

The P5122 Probe is a high voltage probe with 100X
attenuation and a compensation range of 10 to 25 pF.
The probe is designed for use with Tektronix TPS2000
and THS3000 series

fl

oating oscilloscopes with high

input impedance (1 M

). The probe has no user- or

Tektronix-serviceable parts.

WARNING.

The reference contact ring (see below) is

always the same voltage as the reference lead. To avoid
arc

fl

ash, always ensure the insulation sleeve or hook

tip is installed to cover the reference contact ring.

Connecting the Probe to the Oscilloscope

Connect the probe as shown in the illustrations below.

Compensating the Probe

Due to variations in oscilloscope input characteristics,
the low-frequency compensation of the probe may need
adjustment after moving the probe from one oscilloscope
channel to another.
If a 1 kHz calibrated square wave displayed at 1 ms/division
shows signi

fi

cant differences between the leading and

trailing edges, perform the following steps to optimize
low-frequency compensation:

1.

Connect the probe to the oscilloscope channel that you
plan to use for your measurements.

2.

Connect the probe to the probe compensation output
terminals on the oscilloscope front panel.

WARNING.

To avoid electric shock, only connect to the

Probe Comp signal on the oscilloscope when making
this adjustment.

3.

Depress the locking pin inside the protective sleeve.

4.

Turn the sleeve and the cable end until the adjustment
is visible through the opening.

5.

Push AUTOSET or otherwise adjust your oscilloscope
to display a stable waveform.

6.

Adjust the trimmer in the probe until you see a perfectly

fl

at-top square wave on the display. (See illustration.)

WARNING.

To avoid electric shock, only use the

insulated adjustment tool when making compensation
adjustments.

7.

Turn the sleeve back and lock the sleeve in place.

WARNING.

To avoid electric shock, ensure that the

sleeve is locked in the closed position after making
compensation adjustments.

Connecting the Probe to the Circuit

Use the standard accessories included with the probe to
connect to your circuit. (See Table 1.)

WARNING.

To avoid electric shock when using the probe

or accessories, keep

fi

ngers behind the

fi

nger guard of

probe body & accessories (as shown).

To reduce risk of shock, when using the probe on

fl

oating

measurements, ensure the reference lead accessories are
fully mated before connecting the probe to the circuit
under test.

Standard Accessories

The standard accessories included with the probe can be
reordered as a kit. The kit includes one of each item shown
below. Order Tektronix part number 020-3046-00.

Table 1: Probe accessories kit

Item

Description
Hook tip

1. Press onto
probe tip
2. Squeeze the

hook tip to extend
the conductive
hook
3. Clamp hook
onto conductor

Reference lead,
with hook clip

1. Secure lead to

probe head ground
pin
2. Connect hook
clip to circuit
reference

Reference
lead, with
mini-alligator clip

1. Secure lead to
probe head ground

pin
2. Squeeze clip
and connect to
circuit reference

CAUTION.

This accessory is rated 30 VAC

maximum. Do not use for MAINS connection.

Ground spring
for probe tip

1. Press onto
probe tip
2. Probe circuits
with nearby
reference test
points

Adjustment tool

Use only this
insulated tool for

compensation
adjustments

P5122

200 MHz 100X High Voltage Probe

Instructions

x

*P071272003*

071-2720-03