Appendix G: The automated test equipment (ATE) interface
bool SetReceiverOff(classname);
Description: Turns the receiver off in the active controller. Returning True =
Successful.
CAUTION.
Ensure laser power is reduced to zero before running this command,
otherwise the photoreceiver could be damaged.
NOTE.
To turn off lasers, click the LRCP software Laser Emission button to
Off
,
or press the OM2012 front-panel
Power
button to power off the instrument.
Controller Types: 4006, 4106
Example:
SetReceiverOff(Obj);
Returns: ans = true
bool SetReceiverOn(classname);
Description: Turns the receiver on in the active controller. Returning True =
Successful.
Controller Types: 4006, 4106
Example:
SetReceiverOn(Obj);
Returns: ans = true
void SetDesiredFrequency(double desiredFrequency);
Description: Sets the channel and
fi
ne tune frequency of the currently selected
laser to the speci
fi
ed frequency.
Controller Types: 2210, 2012
void TogglePolarization(classname);
Description: Toggles the polarization state by moving both
fi
lters to the
opposite position.
Controller Types: 2210
Example:
TogglePolarization(Obj);
The OUI4000 ATE interface
The Automated Test Equipment (ATE) interface exposes the OUI4006 user
interface functionality through one of two types of Windows Communication
Foundation (WCF) services. The WCF services (referred to as advanced and
basic) are available on port 9200 in the machine that is running the OUI4006.
NOTE.
Because of an optimization in the OUI4006 application, when you attempt
to read variables from MATLAB through this interface you must have the related
plots displayed in the OUI4006 application or the values will not be calculated.
OM4000D Series Coherent Lightwave Signal Analyzer
159
Summary of Contents for OM4006D
Page 2: ......
Page 6: ......
Page 22: ...Compliance information xvi OM4000D Series Coherent Lightwave Signal Analyzer...
Page 24: ...Preface xviii OM4000D Series Coherent Lightwave Signal Analyzer...
Page 100: ...Taking measurements 76 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 146: ...Appendix D Automatic receiver deskew 122 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 202: ...Appendix H Cleaning and maintenance 178 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 205: ...Index W Waveform averaging 48 OM4000D Series Coherent Lightwave Signal Analyzer 181...