Appendix F: Con
fi
guring two Tektronix 70000 series oscilloscopes
Turn off Channels 2 and 4 when trigger is setup.
Set the oscilloscope for 100 Gs/s operation on channels 1 and 3.
Select the maximum corrected bandwidth (not the HW setting).
OUI settings for 2-oscilloscope operation
1.
On the OUI, click Connect on the Setup Tab.
2.
Click
Enable Slave Connection
to enable the selection of a second IP
address. Find the two oscilloscopes and decide which one is the Master. The
Master oscilloscope is the one receiving the external trigger. The Slave scope
is the one triggering on the sync board output only.
If the OUI Scope Connection Dialog box reports “0 Scopes Found,” you will
have to type in the IP address manually. This happens when running over a
VPN or when network policies prevent the IP broadcast.
After connection, the drop-down boxes can be populated. You can select
channels from both the Master and Slave instruments.
150
OM4000D Series Coherent Lightwave Signal Analyzer