Commands listed in alphabetical order
Examples
CH4:PROBEFUNC:EXTUNITS "Pascals"
sets the unit of measurement for the
Channel 4 external attenuator.
CH2:PROBEFUNC:EXTUNITS?
might return
:CH2:PROBEFUNC:EXTUNITS
"Pascals"
, indicating that the Channel 2 external attenuator units of
measurement are Pascals.
CH<x>:PROBEFunc:EXTUnits:STATE
This command sets or queries the custom units enable state for the speci
fi
ed
channel. The channel is speci
fi
ed by x.
Group
Vertical
Syntax
CH<x>:PROBEFunc:EXTUnits:STATE {ON|OFF|<NR1>}
Arguments
OFF
argument turns off external units.
ON
argument turns on external units.
<NR1>
= 0 turns off external units; any other value turns on external units.
Examples
CH2:PROBEFunc:EXTUnits:STATE ON
turns on external units.
CH2:PROBEFunc:EXTUnits:STATE?
might return
:CH2:PROBEFUNC:EXTUNITS:STATE 0
indicating that external units are
off for the speci
fi
ed channel.
CH<x>:SCAle
This command sets or returns the vertical scale for the speci
fi
ed analog channel.
The channel is speci
fi
ed by x.
Group
Vertical
Syntax
CH<x>:SCAle
<NR3>
Arguments
<NR3>
is the vertical scale for the speci
fi
ed analog channel.
MSO54, MSO56, MSO58, MSO58LP Programmer
2-203
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...