Commands listed in alphabetical order
Conditions
Requires option SR-AERO.
Group
Bus
Syntax
BUS:B<x>:MIL1553B:LOWTHRESHold <NR3>
Related Commands
Arguments
<NR3>
is the MIL-STD-1553 lower threshold for the speci
fi
ed bus.
Examples
BUS:B3:MIL1553B:LOWTHRESHOLD -200e-3
sets the Bus 3 MIL-STD-1553
lower threshold to -200 mV.
BUS:B2:MIL1553B:LOWTHRESHOLD?
might return
:BUS:B2:MIL1553B:LOWTHRESHOLD -500.0000E-3
indicating the Bus 2
MIL-STD-1553 lower threshold is set to -500 mV.
BUS:B<x>:MIL1553B:POLarity
This command sets or queries the source polarity for the speci
fi
ed MIL-STD-1553
bus. The bus is speci
fi
ed by x.
Conditions
Requires option SR-AERO.
Group
Bus
Syntax
BUS:B<x>:MIL1553B:POLarity
{NORMal|INVERTed}
BUS:B<x>:MIL1553B:POLarity?
Arguments
NORMal
speci
fi
es normal polarity.
INVERTed
speci
fi
es inverted polarity.
Examples
BUS:B3:MIL1553B:POLARITY INVERTED
sets Bus 3 MIL-STD-1553 polarity
to inverted.
BUS:B2:MIL1553B:POLARITY?
might return
:BUS:B2:MIL1553B:POLARITY
NORMAL
, indicating that the Bus 2 MIL-STD-1553 polarity is set to normal.
2-156
MSO54, MSO56, MSO58, MSO58LP Programmer
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...