![Tektronix MDO4000C Series User Manual Download Page 79](http://html1.mh-extra.com/html/tektronix/mdo4000c-series/mdo4000c-series_user-manual_1079047079.webp)
Triggering on buses
You can use your oscilloscope to trigger on multiple data buses, if you have the
appropriate application module installed. The MDO4000C Series with Option
MDO4MSO can trigger on parallel buses without an application module. The
oscilloscope can display both the physical layer (as analog waveforms) and the
protocol level information (as digital and symbolic waveforms).
To set up the bus trigger:
1.
If you have not already defined your bus using the front panel
B1
,
B2
, or
B3
buttons, do so now.
2.
Push Trigger
Menu
.
3.
Push
Type
.
4.
Turn multipurpose knob
a
to scroll through the trigger type side menu until
you select
Bus
.
5.
Push
Source Bus
and use the Source Bus side menu to select the bus that you
want to trigger on.
6.
Push
Trigger On
and select the desired trigger on feature from the side
menu.
Trigger setup
MDO4000C Series Oscilloscope User Manual
57
Preliminary