5.
Push
Signal Path
on the lower menu.
Utility Page
Calibration
Signal Path
Pass
Factory
Pass
4
5
6.
Push
OK Compensate Signal Paths
on the resulting side menu.
OK
Compensate
Signal Paths
6
The calibration will take approximately 10 minutes to complete.
7.
After calibration, verify that the status indicator on the lower menu displays
Pass
.
If it does not, then recalibrate the instrument or have the instrument serviced
by qualified service personnel.
Service personnel use the factory calibration functions to calibrate the
internal voltage references of the oscilloscope using external sources. Contact
your Tektronix field office or representative for assistance with factory
calibration.
Utility Page
Calibration
Signal Path
Pass
Factory
Pass
7
7
NOTE.
Signal Path Compensation does not include calibration to the probe tip.
Installation
MDO4000C Series Oscilloscope User Manual
25
Preliminary