Installation
Signal Path Compensation
Signal Path Compensation (SPC) corrects for DC inaccuracies caused by temperature variations and/or long-term drift.
Run the compensation whenever the ambient temperature has changed by more than 10 °C (18 °F) or once a week
if you use vertical settings of 5 mV/division or less. Failure to do so may result in the instrument not meeting warranted
performance levels at those volts/div settings.
Signal Path Compensation for Time and Frequency Domains
To compensate the signal path:
1.
Warm up the oscilloscope for at least
20 minutes. Remove all input signals
(probes and cables) from channel inputs.
Input signals with AC components adversely
affect SPC.
2.
Push
Utility
.
3.
Push
Utility Page
.
Utility
Page
4.
Turn
Multipurpose a
and select
Calibration
.
Calibration
5.
Push
Signal Path
on the lower menu.
Utility
Page
Calibration
Signal
Path
Pass
Factory
Pass
MDO3000 Series Oscilloscopes User Manual
21