Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-82
2470-901-01 Rev. A /
May
2019
Example
file_num = file.open("testfile.txt",
file.MODE_WRITE)
if file_num != nil then
file.write(file_num, "This is my test file")
file.close(file_num)
end
Opens file
testfile.txt
for writing. If no
errors were found while opening, writes
This
is my test file
and closes the file.
Also see
(on page 14-77)
format.asciiprecision
This attribute sets the precision (number of digits) for all numbers returned in the ASCII format.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
No
Restore configuration
Instrument reset
Power cycle
Configuration script
0 (Automatic)
Usage
precision
= format.asciiprecision
format.asciiprecision =
precision
precision
A number representing the number of digits to be printed for numbers printed with
the
print()
,
printbuffer()
, and
printnumber()
functions; must be a number
from
1
to
16
; set to
0
to have the instrument select the precision automatically based
on the number that is being formatted
Details
This attribute specifies the precision (number of digits) for numeric data printed with the
print()
,
printbuffer()
, and
printnumber()
functions. The
format.asciiprecision
attribute is only
used with the ASCII format. The precision value must be a number from 0 to 16.
Note that the precision is the number of significant digits printed. There is always one digit to the left
of the decimal point; be sure to include this digit when setting the precision.
Example
format.asciiprecision = 10
x = 2.54
printnumber(x)
format.asciiprecision = 3
printnumber(x)
Output:
2.54000
2.54e+00
Also see
(on page 14-83)
(on page 14-84)
(on page 14-99)
(on page 14-100)
(on page 14-103)