Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-144
2470-901-01 Rev. A /
May
2019
:TRIGger:BLENder<n>:MODE
This command selects whether the blender performs OR operations or AND operations.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Trigger blender clear
Save settings
AND
Usage
:TRIGger:BLENder<n>:MODE <operation>
:TRIGger:BLENder<n>:MODE?
<n>
The blender number (up to two)
<operation>
The type of operation:
OR
AND
Details
This command selects whether the blender waits for any one event (
OR
) or waits for all selected
events (
AND
) before signaling an output event.
Example 1
:DIG:LINE3:MODE TRIG, IN
:DIG:LINE5:MODE TRIG, IN
:TRIG:BLEN1:MODE OR
:TRIG:BLEN1:STIM1 DIG3
:TRIG:BLEN1:STIM2 DIG5
Set digital I/O lines 3 and 5 as trigger in lines. Generate a
trigger blender 1 event when a digital I/O trigger happens on
line 3 or 5.
Also see
:TRIGger:BLENder<n>:STIMulus<m>
(on page 12-145)
:TRIGger:BLENder<n>:OVERrun?
This command indicates whether or not an event was ignored because of the event detector state.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:BLENder<n>:OVERrun?
<n>
The blender number (up to two)
Details
Indicates if an event was ignored because the event detector was already in the detected state when
the event occurred. This is an indication of the state of the event detector that is built into the event
blender itself.