Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-126
2470-901-01 Rev. A /
May
2019
:TRACe:LOG:STATe
This command indicates if information events are logged when the specified reading buffer is at 0% or 100%
filled.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
defbuffer1: ON (1)
defbuffer2: ON (1)
User-created buffer: OFF (0)
Usage
:TRACe:LOG:STATe <logState>
:TRACe:LOG:STATe <logState>, "<bufferName>"
:TRACe:LOG:STATe?
:TRACe:LOG:STATe? "<bufferName>"
<
logState
>
Do not log information events:
OFF
or
0
Log information events:
ON
or
1
<
bufferNam
e>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
Details
If this is set to on, when the reading buffer is cleared (0% filled) or full (100% filled), an event is
logged in the event log. If this is set to off, reading buffer status is not reported in the event log.
Example
TRACe:LOG:STATe?
Query the log state of
defbuffer1
.
Output:
1
Indicates that the log state is on.