2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-47
[:SENSe[1]]:<function>:DELay:USER<n>
This command sets a user-defined delay that you can use in the trigger model.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Function change
Save settings
Measure configuration list
0
Usage
[:SENSe[1]]:<function>:DELay:USER<n> <delayTime>
[:SENSe[1]]:<function>:DELay:USER<n> <DEF|MIN|MAX>
[:SENSe[1]]:<function>:DELay:USER<n>?
[:SENSe[1]]:<function>:DELay:USER<n>? <DEF|MIN|MAX>
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
<n>
The user delay to which this time applies (1 to 5)
<delayTime>
The delay (0 for no delay, or 167 ns to 10 ks)
<DEF|MIN|MAX>
The
DEFault
,
MINimum
, or
MAXimum
value
Details
To use this command in a trigger model, assign the delay to the dynamic delay block using the
corresponding
MEAS<n>
parameter that matches the delay number specified here (see the Example
below).
The delay is specific to the selected function.
Example
:CURRent:DELay:USER1 0.2
:TRIGger:BLOCk:DELay:DYNamic 6, MEAS1
Set user delay 1 to 0.2 s for current
measurements. Set trigger block 6 to be a
dynamic delay that is set to user delay 1 for
the function being measured.
Also see