Section 3: Instrument description
2470 High Voltage SourceMeter Instrument Reference Manual
3-42
2470-901-01 Rev. A /
May
2019
From the front panel:
1.
Press the
MENU
key.
2.
Under System, select
Settings
.
3.
Select the button next to Reading Format.
4.
Select the reading format (
Prefix
or
Exponent
).
This setting takes effect immediately.
Over a remote interface:
•
(on page 12-31)
•
(on page 14-68)
Customizing a message for the USER swipe screen
You can customize the message that is displayed on the USER swipe screen.
You must use a remote interface to customize the USER swipe screen.
Creating a message
When you create the message, you can send text that will be used on the top and bottom lines of the
USER swipe screen. The top line allows up to 20 characters and the bottom line allows up to 32
characters.
The examples shown here switch the display to the USER swipe screen, set the first line to read
"Test in process," and the second line to display "Do not disturb."
Using SCPI commands:
Send the commands:
DISPlay:SCReen SWIPE_USER
DISPlay:USER1:TEXT "Test in process"
DISPlay:USER2:TEXT "Do not disturb"
Using TSP commands:
Send the commands:
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Test in process")
display.settext(display.TEXT2, "Do not disturb")