Model 2461 Calibration and Adjustment Manual
Section 5: Command reference
2461-905-01 Rev. A / February 2016
5-7
:CALibration:DIGitize:ADJust:SENSe
This command adjusts the Model 2461 digitizer function.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:CALibration:DIGitize:ADJust:SENSe
Details
This command must be sent after all other
:CALibration:ADJust
commands for all ranges have
been completed successfully.
The Model 2461 does not source any output and there is no need to connect external equipment
during this adjustment step.
This command initiates an internal operation to adjust digitizer circuits to match the readings of
integration analog-to-digital circuits.
This command requires about 10 seconds to complete.
Example
-- Perform all other adjustment steps
first.
:CAL:DIG:ADJ:SENS; *OPC?
The
*opc?
query returns "1" when the
adjustment command is finished. This command
may require a longer timeout value than most
commands.
Also see
(on page 5-3)
(on page 5-5)