Section 2: Measuring dc voltage with high accuracy
Model DMM7510 7½ Digit Multimeter Application Manual
2-2
DMM7510-904-01 Rev. D March 2021
Make sure power to the instrument is off and make the following connections:
•
Connect the test leads to the INPUT HI and LO terminals.
•
Connect the INPUT HI connection to one of the DUT leads.
•
Connect the INPUT LO connection to the other DUT lead.
The physical connections for the front and rear panels are shown in the following figures.
Figure 1: Front-panel connections for high-accuracy dc voltage measurements
Figure 2: Rear-panel connections for high-accuracy dc voltage measurements