In this section:
Replacing the boards................................................................ 9-1
How to clean the Model 8010 Test Fixture ............................... 9-1
Connector and socket cleaning ................................................ 9-2
Replacing the boards
The Model 8010 is shipped with two 8010-DTB Device Test Boards installed. The boards include
connections for TO-247 and axial-lead devices.
The Model 8010 is also shipped with an 8010-CTB Customizable Test Board, which allows you to
add your own connections.
In addition, you can purchase the 8010-DTB-220 Device Test Board for use with three-terminal
TO-220 or TO-247 devices. This board is limited to 1000 V. The 8010-DTB-CT Curve Tracer Adapter
Test Board is also available, and allows you to use the same socket style installed in Tektronix Series
Curve Tracers.
You can replace any of these boards as needed. For replacement instructions, see
(on page 2-3).
How to clean the Model 8010 Test Fixture
To prevent contamination that may degrade performance, use care when handling or servicing the
test fixture.
To avoid damage to the test fixture, do not use any abrasive or chemical cleaning agents.
To clean the Model 8010:
1. Remove power from all instruments that are connected to the test fixture.
2. Remove loose dust on the outside and inside of the test fixture with a lint-free cloth.
3. Use a soft cloth or swap dampened with water to clean the test fixture. Use an aqueous solution
of 75% isopropyl alcohol for more efficient cleaning. Avoid getting moisture inside the test fixture
during cleaning. Use only enough cleaning solution to dampen the cloth or swab.
If a board becomes contaminated and operation is affected, the board should be returned to the
factory for proper cleaning and servicing.
Section 9
Maintenance